Eppendorf Innova 40R Shaker User Manual
Page 11
Product description
New Brunswick Innova 40/40R Shaker
English (EN)
11
Temperature control
Ambient temperature is measured at 1
M from the exterior of the unit.
• Innova
40R provides temperature control from 15
°C below ambient temperature to 80
°C with a
minimum setpoint of 4
°C.
• Innova
40 provides temperature control from 5
°C above ambient temperature to 80
°C.
Both these ranges depend on relative humidity and other ambient factors, as well as the options installed
on the device.
Safety
• Drive Interrupt shuts off power to shaker when lid opens.
• Acceleration/deceleration circuit prevents sudden starts and stops, minimizing both splashing and
mechanical damage.
• Independent mechanical sensing switch also shuts the motor off in unbalanced condition.
Alarms
Innova 40/40R Shaker is equipped with visual and audible alarms that alert the user to the following
conditions:
• The end of a timed run.
• Deviations from speed setpoint.
• Deviations from temperature setpoint.
• Power failure.
• Lid is open.
Platforms
To accommodate customer needs, a wide variety of platforms can be used with the Innova 40/40R Shaker
(see
• Universal platforms are the most flexible, providing hole patterns for flask clamps, test tube racks and
other accessories.
• Dedicated platforms are supplied with 1 size flask clamps attached.
Flasks/Tubes
Erlenmeyer flasks (up to 3
L), and a wide variety of tubes and plates can be accommodated using the New
Brunswick shaker accessories (see
Other accessories
Test tube racks, microplate holders, and test tube rack holders as well as sticky pads and stick tape are also
available (a universal platform is needed for all test tube racks and holders) (see