Maintenance
Eppendorf ThermoMixer

®

C

English (EN)

7

Maintenance

7.1

Setting service intervals

The ThermoMixer C offers the option of activating a reminder that the device needs to be
serviced. To set a service interval, proceed as follows:

1. Under

Menu

>

Settings

> select the

Service

menu item. Confirm with the

menu/enter

key.

2. Select a service interval with the menu arrow keys (after 500, 1 000 or 2 000 operating

hours).
To switch off the notification, select

No notification

.

When the specified operating hours have been reached, a message appears. Contact your
local Eppendorf partner. The contact addresses can be found online at
www.eppendorf.com/worldwide.

7.2

Cleaning

Clean the housing of the ThermoMixer C and the thermoblocks regularly.

DANGER! Electric shock as a result of penetration of liquid.

Switch off the device and disconnect the power plug before starting cleaning
or disinfection work.

Do not allow any liquids to penetrate the inside of the housing.

Use closed tubes and closed plates.

Do not spray clean/spray disinfect the housing.

Only plug the device back in if it is completely dry, both inside and outside.

NOTICE! Damage from the use of aggressive chemicals.

Do not use any aggressive chemicals on the device or its accessories, such as
strong and weak bases, strong acids, acetone, formaldehyde, halogenated
hydrocarbons or phenol.

If the device has been contaminated by aggressive chemicals, immediately
clean it by means of a mild cleaning agent.

NOTICE! Corrosion from aggressive cleaning agents and disinfectants.

Do not use corrosive cleaning agents, aggressive solvents or abrasive
polishes.

Do not use lab cleaners with sodium hypochlorite.