Safety
Eppendorf ThermoMixer

®

C

English (EN)

WARNING! Risk of injury due to incorrect consumables.

• Poorly fitting tubes or plates can become detached from the thermoblock.

• Glass tubes can smash.

Only use the thermoblocks with the consumables designed for them.

Never use tubes made of glass or other fragile material.

WARNING! Contamination due to opening seals of consumables.
In the following cases, the seals of tubes can spring open. Sample material can
escape.

• high vapor pressure of the content

• improperly sealed cover

• damaged sealing lip

• improperly fastened foil

Always check that consumables have been sealed tightly before use.

WARNING! Injury from sample material being thrown out.
Sample material can be thrown out of open, improperly sealed or unstable tubes
and plates.

Only mix in closed tubes and closed plates.

Observe the nationally prescribed safety environment when working with
hazardous, toxic and pathogenic samples. Pay particular attention to personal
protective equipment (gloves, clothing, goggles etc.), extraction, and the
safety class of the lab.

CAUTION! Poor safety due to incorrect accessories and spare parts.
The use of accessories and spare parts other than those recommended by
Eppendorf may impair the safety, functioning and precision of the device.
Eppendorf cannot be held liable or accept any liability for damage resulting from
the use of incorrect or non-recommended accessories and spare parts, or from
the improper use of such equipment.

Only use accessories and original spare parts recommended by Eppendorf.

CAUTION! Crush hazard due to moving parts.

Do not replace any consumables during the mixing process.

Do not remove the thermoblock during the mixing process.

Put on the ThermoTop or Lid prior to the mixing process.