beautypg.com

Eppendorf ThermoMixer C User Manual

Page 18

background image

18

Operation
Eppendorf ThermoMixer

®

C

English (EN)

Abb. 5-2:Display ThermoMixer C

Fig. 5-2:

Display ThermoMixer C

1

Program number

2

Program name

3

Program levels (step 1 to step 4)
*: current step

4

Device status

Device is performing mixing/

temperature control.

Mixing process interrupted,

temperature control to be continued.

5

ThermoTop

ThermoTop has been attached.

To prevent the formation of condensate,
the device heats up the ThermoTop,
before the temperature of the
thermoblock is controlled.

6

Interval Mix

Interval Mix has been activated for

the current step.

7

Speaker

Speaker switched on.

Speaker switched off.

8

Key lock

Key lock activated: parameters cannot

be changed.

No key lock.

9

Time mode

Time Control

Time counting begins

immediately.

Temp Control

Time counting begins

when the set temperature has been
reached.

10 Mixing frequency

11 Actual temperature

When the actual temperature flashes on
the display, the device is not in
temperature control mode operation.

12 Set temperature

When the set temperature has been
reached, only one value is displayed.

13 Mixing time

1

2

3

4

5

6

7

8

9

10

11

12

13