Eppendorf ThermoMixer C User Manual
Page 32

32
Operation
Eppendorf ThermoMixer
®
C
English (EN)
5.9.2
Interval Mix: Creating a program level with pause interval
By using the Interval Mix function, you can determine the mixing process to be
interrupted by one or several pauses
within one program level . Interval Mix can only be
specified in programs. For program levels with Interval Mix, the
symbol will appear in
the display
1. Select an empty program space under
Menu
>
Programs
.
2. Set the mixing time, temperature, and mixing frequency with the
time, temp and
speed arrow keys.
Select the mixing time so that it covers the entire duration, including mixing processes
and pauses.
3. Select
Options
. Confirm with the
menu/enter key.
4. Select
Interval Mix
. Confirm with the
menu/enter key.
5. Set the mixing time (before the pause) in the
Mixing time
row using the
time arrow
keys.
6. Set the pause duration in the
Pause
row using the
time arrow keys.
7. In order to save the program, select
Save
. Confirm with the
menu/enter key.
The Interval Mix settings are now saved for the program level.
To program a change between several mixing processes and pauses within a
program level, select a correspondingly longer mixing time for the program
level:
Change between mixing process and pause:
• Mixing time set for program level: 6:00 min
• Interval Mix:
Mixing time
: 1:00 min,
Pause
: 0:30 min
In the program level the device performs the setting 1 min mixing and 0:30 min
pause 4 times.