5.9.2

Interval Mix: Creating a program level with pause interval

By using the Interval Mix function, you can determine the mixing process to be
interrupted by one or several pauses

within one program level . Interval Mix can only be

specified in programs. For program levels with Interval Mix, the

symbol will appear in

the display

1. Select an empty program space under

Menu

>

Programs

.

2. Set the mixing time, temperature, and mixing frequency with the

time, temp and

speed arrow keys.
Select the mixing time so that it covers the entire duration, including mixing processes
and pauses.

3. Select

Options

. Confirm with the

menu/enter key.

4. Select

Interval Mix

. Confirm with the

menu/enter key.

5. Set the mixing time (before the pause) in the

Mixing time

row using the

time arrow

keys.

6. Set the pause duration in the

Pause

row using the

time arrow keys.

7. In order to save the program, select

Save

. Confirm with the

menu/enter key.

The Interval Mix settings are now saved for the program level.

To program a change between several mixing processes and pauses within a
program level, select a correspondingly longer mixing time for the program
level:
Change between mixing process and pause:

• Mixing time set for program level: 6:00 min

• Interval Mix:

Mixing time

: 1:00 min,

Pause

: 0:30 min

In the program level the device performs the setting 1 min mixing and 0:30 min
pause 4 times.