6 temperature control, 1 temperature control with time setting, Temperature control 5.6.1 – Eppendorf ThermoMixer C User Manual
Page 24: Temperature control with time setting

24
Operation
Eppendorf ThermoMixer
®
C
English (EN)
5.6
Temperature control
The ThermoMixer C can be used for temperature control in a range of 15 °C below the
ambient temperature to 100 °C.
5.6.1
Temperature control with time setting
Prerequisites
The time mode is set to
Time Control
1. To switch off the mixing function, use the
speed arrow keys to select the 0 rpm setting
(▼ below 300 rpm or ▲ above 3 000 rpm).
2. Using the
time arrow keys set the temperature control duration.
3. Using the
temp arrow keys set the temperature.
4. To start time counting, press the
start/stop key.
• The symbol flashes on the display.
• The temperature control duration is counted down.
• The display shows the remaining temperature control duration and the actual
temperature/set temperature.
• A signal sounds after the temperature control duration has elapsed.
Functioning principle of the ThermoTop
• In order to prevent the formation of any condensate in a reliable manner, the
device first heats the ThermoTop until it reaches the set temperature. The
tempering of the thermoblock occurs with a delay.
• The temperature sensor of the thermoblock reacts to the temperature of
samples: after inserting samples into a pre-heated thermoblock, the displayed
actual temperature may fall temporarily.
• While the device is tempering, the blue LED of the ThermoTop is flashing.
NOTICE! Damage to electronic components due to condensation.
Condensate can form in the device after it has been moved from a cool
environment to a warmer environment.
After installing the device, wait at least for 3 h. Only then connect the device
to the mains.
• When the actual temperature flashes on the display, the device is not in
temperature control mode operation.
• As soon as the set temperature is changed with the
temp arrow keys, the
device begins to perform temperature control.
• When the set temperature has been reached, the display only shows one
value.