7 mixing, 1 mixing without temperature control, 2 mixing and tempering – Eppendorf ThermoMixer C User Manual
Page 25: Mixing 5.7.1, Mixing without temperature control, Mixing and tempering

Operation
Eppendorf ThermoMixer
®
C
English (EN)
25
5.7
Mixing
5.7.1
Mixing without temperature control
1. In order to switch off the temperature control, use the
temp arrow keys for selecting
the
off
setting (▼ below 1 °C or ▲ above 100 °C).
2. Set the mixing time by means of the
time arrow keys.
3. Set the mixing frequency by means of the
speed arrow keys.
4. In order to start the mixing process, press the
start/stop key.
• The symbol flashes on the display.
• The mixing time is counted down.
• The display shows the remaining mixing time, the actual temperature/set
temperature and the mixing frequency.
5. After the set mixing time has elapsed, the device stops automatically.
• A signal tone sounds.
• The display shows the last used parameters.
5.7.2
Mixing and tempering
1. Set the mixing time with the
time arrow keys.
2. Set the temperature with the
temp arrow keys.
The device immediately starts to perform the temperature control.
3. Set the mixing frequency with the
speed arrow keys.
4. In order to start the mixing process, press the
start/stop key.
• The symbol flashes on the display.
• The mixing time is counted down.
• The display shows the remaining mixing time, the actual temperature/set
temperature and the mixing frequency.
You can select the mixing frequency between 300 rpm and 3 000 rpm. The
mixing frequency can be adjusted in steps of 50 rpm.