SERVICE
26
5.
Close the sliding door.
Figure 3. Door Stop
6.
Working at the back of the mixer, gently slide the door out of the two
notches in the top of the right and left side door tracks (see Figure 3).
Set door aside.
7.
Clean the door tracks.
8.
Place the new door on a clean, flat surface.
9.
Vertically apply masking tape to the slats in order to keep the slats in
line.
10.
Feed the replacement sliding door into the two notches on the top right
and left side door tracks.
11.
Gently glide the door all the way down.
12.
Remove the masking tape.
13.
Test the operation of the door by opening and closing. Make
adjustments, as necessary.
14.
Mount the door stop back onto the right track with the two (2) screws.
15.
Reassemble the top cover and rear access panel.
16.
Restore power to the mixer.
TRACK
DOOR STOP