Troubleshooting gyromixer end user guide 19 – Fluid Management Miller GyroMixer User Manual
Page 23

TROUBLESHOOTING
GyroMixer End User Guide
19
Mixer will not shut off.
Damaged timer relay.
•
Replace timer relay.
Faulty connection to the potentiometer.
•
Tighten connection to potentiometer.
Damaged potentiometer.
•
Replace potentiometer.
Mixer hums, but will not run.
Insufficient current.
•
Verify mixer is correctly installed to a
dedicated line.
Something caught in mixer.
•
Check for obstructions.
Defective bearing in pulleys four arms or
main drive.
•
Check each bearing for roughness.
Replace if necessary.
Defective motor.
•
Replace motor.
Mixer squeaks at startup, but still
runs.
Belts loose.
•
Check V-belt and poly-V belt
tension. Tighten if necessary.
PROBLEM
POSSIBLE CAUSE
ACTION