beautypg.com

Bacharach Fyrite INSIGHT User Manual

Page 9

background image

Instruction 24-9460

12

Fyrite

®

INSIGHT

Instruction 24-9460

13

Fyrite

®

INSIGHT

3.5.3 Pressure Units

Select to display pressure in inches of water column (inwc), millibar (mb),

Pascals (Pa), hectoPascals (hPa), or millimeter of water column (mmwc) as

follows:

1. Enter the Setup Menu per Section 3.5.1. If necessary, press ESC

until MENU appears above F2.

2. From the SETUP MENU, use the ▲▼ but-

tons to select PRESSURE UNITS. Press
ENTER

to display the Pressure Units Menu.

3. Use the ▲▼ buttons to selecte the desired

pressure unit.

4. Press ENTER to save selection and return

to the SETUP Menu or ESC to revert to the
previous setting.

3.5.4 Clock

The clock settings (12 hr. AM/PM or 24 hr.) and date format options (MM/

DD/YY or DD/MM/YY) are interdependent. (See DATE FORMAT, Section

3.5.15).

1. Enter the Setup Menu per Section 3.5.1. If necessary, press ESC

until MENU appears above F2.

2. From the SETUP MENU, use the▲▼ but-

tons to select CLOCK. Press ENTER to

display the Clock Setup Screen

3. Use the ▲▼ buttons to enter the current

date and time. Use the the ◄► buttons to

move the cursor across the screen.

4. Press ENTER to save the selection and re-

turn to the SETUP MENU or ESC to revert

to the previous setting.

3.5.5 O

2

Reference

The measured value of CO can be referenced to a specific O2 percentage of

0% to 15%. Set up the reference value as follows:

1. Enter the Setup Menu per Section 3.5.1. If

necessary, press ESC until MENU appears

above F2.

2. From the SETUP MENU, use the ▲▼

buttons to select O2 REFERENCE. Press
ENTER

to display the O2 Reference Screen.

3. Use the ▲▼ buttons to enter the O2 refer-

ence value.

4. Press ENTER to save the selection and re-

turn to the SETUP MENU or ESC to revert

to the previous setting.

3.5.6 Print Pressure

Select whether to print or not print the pressure measurement on the com-

bustion test printout as follows:

1. Enter the Setup Menu per Section 3.5.1. If necessary, press ESC

until MENU appears above F2.

2. From the SETUP MENU, use the ▲▼ but-

tons to select PRINT PRESSURE. Press
ENTER

to display the O2 Reference Screen.

3. Use the ▲▼ buttons to select YES or NO to

print the pressure measurement.

4. Press ENTER to save the selection and re-

turn to the SETUP MENU or ESC to revert

to the previous setting.

3.5.7 Zoom

Combustion test data in the Run/Hold screen can be shown with enlarged

characters to make viewing easier. The operator can set zoom levels to

Standard, 2X or 3X. The Standard zoom setting will display 4 lines of

combustion test data at one time; 2X will display 3 lines of data with en-

larged characters; 3X will display 2 lines of data with enlarged characters.

The operator can scroll through the complete list of measured and calcu-