1.3 Operational Overview

Pressing the PWR button turns the analyzer ON. Note that there is a

warm-up period of 60 seconds that must elapse before the analyzer can be

used.

To perform a combustion test, choose a fuel code that corresponds to the

fuel being burned by the appliance being tested (Section 4.3), and then

press the RUN/HOLD button to place the analyzer into its Run Mode.

Begin testing by inserting the analyzer’s probe tube into the flue-gas

stream of the appliance under test. The analyzer will monitor the flue

gas and display measured and calculated values that are relative to the

combustion process. These values are dis played on the analyzer’s LCD, and

are chosen for display by pressing the In cre ment () and Dec re ment ()

but tons.

Pressing the RUN/HOLD button during a test will freeze all measured and

calculated values at their current levels. Pressing the RUN/HOLD button

again resumes testing.

A backlight enables an operator to read the display in dimly-lit areas.

Pressing the PWR button turns the backlight ON and OFF.

An optional power saver function will shut the analyzer OFF after a set

time period of inactivity. This feature is dis abled if the CO value is greater

than 50 ppm or the O

2

value is less than 16.0%.

Press the PWR button for at least two seconds to power the analyzer

OFF. Note that there is a 5 second delay before the analyzer actually

turns OFF, during which time the unit can be kept ON by pressing the
RUN/HOLD

button. There is a CO purge feature that keeps the analyzer’s

pump running until the measured CO level drops below 50 ppm.

2.0 TECHNICAL CHARACTERISTICS

The Analyzer measures & displays:

Primary/Ambient Air Temperature

-4 to 600 °F (-20 to 316°C)

Oxygen

0.1 to 20.9% O

2

Stack Temperature

-4 to 1202 °F (–20 to 650 °C)

Carbon Monoxide

0 to 2000 ppm CO

Differential Pressure/Draft

± 100 inwc

The Analyzer computes & displays:

Combustion Efficiency

0.1 to 100%

Carbon Dioxide

0.1 to a fuel dependent
maximum value in %

CO referenced to %O

2

0 to 9,999 ppm

Excess Air

0 to 250%

Fuel Selection

• Natural Gas
• Oil #2
• Oil #4

• Oil #6

• Propane
• Kerosene
• Wood
• Coal

Normal Operating Conditions:

Temperature

• Analyzer
• Probe

32 to 104 °F (0 to 40 °C)
1,000 °F max (538 °C) at
5" insertion

Humidity

15 to 90% RH, Non-Condensing

Dimensions

Weight

7.9 in (20.1 cm) Height
3.6 in (9.1 cm) Width
2.3 in (5.8 cm) Depth

1lb. (0.45 kg)