beautypg.com

10 status screen, 9 diagnostics screen – Bacharach Fyrite INSIGHT User Manual

Page 29

background image

Instruction 24-9460

52

Fyrite

®

INSIGHT

Instruction 24-9460

53

Fyrite

®

INSIGHT

Main Diagnostics - Lists the current status of the Stack and Air Ther-

mocouple channels, Reference Temperature channel, Pressure channel,

CO channel, O

2

channel and Battery.

O2 Sensor Life - Displays the approximate remaining life of the Oxygen

sensor.

Fresh Air Diagnostics - Fresh air diagnostics will cycle the instrument

through the 60 second warm-up to check on the status of the sensors. The

probe must be in fresh air to perform a valid check.

Access the DIAGNOSTIC menu as follows:

1. Display the MAIN MENU by pressing the MENU (F2) button. If neces-

sary, press ESC until MENU appears above F2.

2. Use the ▲▼ buttons to highlight DIAGNOSTICS and then press ENT to

display the DIAGNOSTIC MENU.

3. Use the ▲▼ buttons to highlight the desired diagnostic topic and then

press ENT to display the information under that topic.

4. Press the ESC key to exit back to the DIAGNOSTIC MENU or the

MENU (F2)

key to exit back to the MAIN MENU.

6.10 Status Screen

The Status Screen provides a quick reference to key items when trouble-

shooting:

• Version
• ADC Version
• Boot Version
• Model
• SN
Access the Status menu as follows:

1. Display the MAIN MENU by pressing the MENU (F2) button. If neces-

sary, press ESC until MENU appears above F2.

2. Use the ▲▼ buttons to highlight STATUS and then press ENT to dis-

play the DEVICE STATUS screen.

3. Press the ESC key to exit back to the MAIN MENU.

Low Sensor- O2 or CO sensor outputs were low but still usable.

Sensor(s) may need to be replaced in the near future. Message will indi-

cate which sensor(s) were in warning.

Warmup Sensor Error -

• CO sensor was not zeroed at warmup because of high output. Run

instrument on fresh air then restart instrument to re-zero sensor. If

message persists, CO sensor may need to be replaced.

• Stack or Air temperature channel is measuring temperature outside

the range of -4 to 212 ˚F at startup. Make sure that the Stack and

Air thermocouples are sampling ambient room air within the tem-

perature range at startup.

• Pressure sensor is measuring pressure outside the range of ± 3 inwc

at startup. Ensure that the analyzer is sampling atmospheric pres-

sure and restart the instrument.

• The analyzer was turned on with the probe sampling flue gas. Move

the probe to fresh air and restart the instrument.

• Message will indicate which channel is in error.

Low Battery - Battery voltage is low. Replace the batteries.

XXXX - Occurs in the number fi elds of sensors that have achieved over-

range condition.

**** - Occurs in the number fi elds of sensors and the calculated values

that depend on the sensors that were in error coming out or warmup.

- - - - Occurs in the number fields of calculated values when Oxygen is

above 16.0%.

6.9 Diagnostics Screen

The DIAGNOSTICS menu provides information regarding the operation of

the analyzer. Information includes the following:

Time Meters - Displays the run time of the analyzer and sample pump in

hours of operation.