Change the order in which data is displayed as follows:

1. Enter the SETUP MENU per Section 3.5.1. If necessary, press ESC

until MENU appears above F2.

2. Use the ▲▼ buttons to select RUN/HOLD

FORMAT. Press ENTER to display the

Run/Hold Format options.

3. Use the ▲▼ buttons to select EDIT FOR-

MAT. Press ENTER to display the current

format.

4. Change data for a particular location by

fi rst using the ▲▼ buttons to select the loca-

tion. Press ENTER and the cursor will start

to fl ash.

5. Use the ▲▼ buttons to scroll through and

select the desired data to appear in that

location. Press ENTER to save the selection.

6. Change the data displayed at other locations

by repeating steps 4 and 5

7. When fi nished, use the ▲▼ buttons to select EDIT COMPLETE,

located at the bottom of the list. Press ENTER to save the new dis-

play format and return to the Run/Hold Format options.

Reset Display format back to factory default settings as follows:

1. Enter the SETUP MENU per Section 3.5.1.

If necessary, press ESC until MENU ap-

pears above F2.

2. Use the ▲▼ buttons to select RESET FOR-

MAT. Press ENTER to display the Reset

Format option.

3. Use the ▲▼ buttons to select YES. Press

ENTER

to reset the format and return to the

Run/Hold format options.

3.5.10 Language Selection

Information on the display screen can be shown in English, French, or

Spanish.

Select the desired language as follows:

1. Enter the SETUP MENU per Section 3.5.1. If necessary, press ESC

until MENU appears above F2.

2. Use the ▲▼ buttons to select LANGUAGE

SELECTION. Press ENTER to display the

Language Selection Menu.

3. Use the ▲▼ buttons to select the desired

language. Press ENTER to activate the

selection and return to the SETUP MENU.

3.5.11 Button Sound

The audible sound used to signal when a button is pressed can be turned

OFF and ON as follows:

1. Enter the SETUP MENU per Section 3.5.1. If necessary, press ESC

until MENU appears above F2.

2. Use the ▲▼ buttons to select BUTTON

SOUND. Press ENTER to display the Button

Sound Menu.

3. Use the ▲▼ buttons to select YES or NO

button sound.

4. Press ENTER to save the selection and re-

turn to the SETUP MENU or ESC to revert

to the previous setting.

3.5.12 CAL Reminder Period

The analyzer can be set to indicate a calibration reminder during the 60

second warm-up. Calibration reminders can be preset to occur never, 6,

8, 10, 12, or 15 months after the last calibration. When the preset period

is exceeded the instrument will display the reminder, and how long since

the sensors were last calibrated. If a calibration reminder is displayed the

operator can press the RUN/HOLD key to move to the Run/Hold Screen for

normal operation. Regular calibration periods of 6 months to 1 year are

recommended.