beautypg.com

0 operation 4.1 operating tips, 2 analyzer power on & warm up, 3 fuel selection – Bacharach Fyrite INSIGHT User Manual

Page 13

background image

Instruction 24-9460

20

Fyrite

®

INSIGHT

Instruction 24-9460

21

Fyrite

®

INSIGHT

4.0 OPERATION

4.1 Operating Tips

When an analyzer is brought in from a cold vehicle, let it warm up

slowly to minimize condensation. Temperatures below freezing will

not damage the analyzer; however, bringing a cold analyzer into a

warm, humid environment may cause condensate to form inside the

case.

Caution

: Although the analyzer itself is not damaged by an ex-

tremely cold environment, the electrochemical sensors may be

damaged. The O

2

sensor's electrolyte will freeze at approximately

-20 °F and the CO sensor's at approximately -90 °F. If the analyzer

is exposed to an extremely cold condition, it is strongly suggested

that the sensor housings be examined for hairline cracks. Be aware

that a leaking sensor can cause chemical burns to the skin and

possibly damage the PCB assemblies.

Ensure that the analyzer is sampling fresh air when turned ON. Pulling

a stack-gas sample through the analyzer during its warm-up period will

not damage the analyzer, but it will result in incorrect sensor readings,

and may result in sensor error messages appearing after the warm-up

cycle completes.

Note that flue-gas condensate is acidic and very corrosive. It is important

not to allow the analyzer's internal components to come in contact with

condensate for long periods of time.

Before each use, inspect the filter element of the water-trap / filter as-

sembly. Replace the filter if it looks dirty.

When sampling flue-gas, keep the analyzer above the water-trap, and keep

the trap in a vertical position. This will maximize the effectiveness of the

trap and keep liquid from being drawn directly into the analyzer.

When liquid condensate is seen inside the water trap, empty the trap

before it becomes full.

When storing the analyzer, it's a good idea to empty the water trap and

leave it open to further dry it out.

Calibrate the analyzer every 6 months - 1 year to assure its accuracy.

4.2 Analyzer Power On & Warm Up

1. Connect the probe and make sure that the analyzer is properly set

up per Sec tion 3.

IMPORTANT!

DO NOT insert probe into stack before powering ON

the analyzer.

2. Place the probe in an area that contains fresh air. This ensures

that the sensors will be properly zeroed during the warm-up cycle.

3. Power ON the analyzer by pressing the

PWR

button for at least 1 second, or until

a single beep is heard. Observe that the

analyzer's firmware version, model and se-

rial numbers are briefly displayed followed

by the Warm Up screen.

4. Wait for the analyzer to count down its 60

second warm-up period; after which, the in-

strument will display the Combustion Test

HOLD screen.

If, however, any errors were detected during

warm-up, the message "ERRORS DETECT-

ED" is displayed along with a list of those

errors. In this example, the O2 sensor is

missing and the T-Stack thermocouple is not

connected. Refer to Section 6.8 for a listing

and possible remedy for the errors displayed.

4.3 Fuel Selection

The top line of Combustion Test HOLD screen shows the fuel currently

selected. If necessary, change the fuel as follows:

1. Display the MAIN MENU by pressing the

MENU (F2)

button. If necessary, press ESC

until MENU appears above F2

2. Use the ▲▼ buttons to select FUEL. Press

ENTER

to display the Fuel Selection screen.

3. Use the ▲▼ buttons to select the desired fuel.

24-8250

V1.00