5.0 CALIBRATION & MAINTENANCE

IMPORTANT:

Before performing any calibration procedure,

ensure that fresh batteries are installed or use the optional

AC power adapter. Also, ensure that the analyzer is at room

temperature and will be sampling fresh air when powered

ON.

IMPORTANT:

To main tain ac cu ra cy as listed in the Techni-

cal Characteristics Sec tion of this manual, the standards

used must be at least four times as accurate as the stated

accuracy of the Fyrite

®

INSIGHT.

5.1 B-Smart Sensors

The Fyrite

®

INSIGHT utilizes Bacharach’s new B-Smart Sensor tech-

nology for its CO sensor. The B-Smart sensor is marked with a 10-digit

calibration code that can be entered in the instrument's memory via the

keypad, or with Fyrite user software.

Benefits of B-Smart Sensors:

New sensors can be installed without applying gas for calibration.

Sensors can be pre-calibrated and installed when needed.

Sensors can be moved from one analyzer to another.

The B-Smart sensors should be calibrated by an authorized Bacharach

Service Center every 6 months to 1 year to assure that the analyzer contin-

ues to meet its published accuracy specifications. The B-Smart sensors,

however, can be calibrated in the field if your facility has the necessary

equipment and qualified personnel to perform the procedures described in

the following sections of this instruction manual.

5.2 Starting a Calibration

Start any calibration by performing the following:

1. With the analyzer power OFF, place the unit in fresh, ambient air;

then power on the analyzer.

After the installation,

• Open Fyrite

®

User Software via the desktop icon.

• Connect the instrument to the PC with a USB cable.

• Power up the unit.

• When first connected, a “Found New Hardware Wizard” screen will be

displayed.

Click on: "Next..."

"Next..."

"Continue anyway"

"Finish"

4.16 Fyrite

®

User Software Description

Fyrite

®

INSIGHT User Software

Set instrument time

Sync instrument with computer's time

Calibration password

Change 4 digit numeric code

Time meters

Provides instrument & pump run time

B-Smart

B-Smart calibration code entry

Saved measurements

Downloads instrument memory to Excel

User/Customer ID

User ID entry

Settings

Change instrument set up parameter

Customer logo set up

Load custom logos for print out. Limited to 384 x 192.

Custom fuel

Bacharach can supply fuel codes for custom fuel. Consult
factory.

Update instrument software

Remote upgrade of instrument software.