Language Dictionary - 4

71

*OPT?

This query requests the dc source to identify any options that are installed. Options are identified by
number A

0 indicates no options are installed.

Query Syntax

*OPT?

Returned Parameters

*RCL

WARNING:

Recalling a previously stored state may place hazardous voltages at the dc source output.

This command restores the dc source to a state that was previously stored in memory with the *SAV
command to the specified location. All states are recalled with the following exceptions:

u

the trigger system is set to the Idle state by an implied ABORt command (this cancels any
uncompleted trigger actions)

u

the calibration function is disabled by setting CAL:STATe to OFF

NOTE:

The device state stored in location 0 is automatically recalled at power turn-on when the
OUTPut:PON:STATe is set to RCL0.

Command Syntax

*RCL

Parameters

0 | 1 | 2 | 3

Example

*RCL 3

Related Commands

*PSC *RST *SAV

*RST

This command resets the to a factory-defined state as defined in the following table. *RST also forces an
ABORt command.

Command Syntax

*RST

Parameters

None

Related Commands

*PSC *SAV