Common Commands

Common commands begin with an * and consist of three letters (command) or three letters and a ?
(query). They are defined by the IEEE 488.2 standard to perform common interface functions. Common
commands and queries are categorized under System, Status, or Trigger functions and are listed at the
end of each group. The dc source responds to the following commands:

Table 4-2. Common Commands Syntax

*CLS

Clear status

*ESE

Standard event status enable

* ESE?

Return standard event status enable

*ESR?

Return event status register

*IDN?

Return instrument identification

*OPC

Enable "operation complete" bit in ESR

*OPC?

Return a "1" when operation complete

*OPT?

Return option number

*PSC

Power-on status clear state set/reset

*PSC?

Return power-on status clear state

*RCL

Recall instrument state

*RST

Reset

*SAV

Save instrument state

*SRE

Set service request enable register

*SRE?

Return service request enable register

*STB?

Return status byte

*TRG

Trigger

*TST?

Perform selftest, then return result

*WAI

Hold off bus until all device commands done

Programming Parameters

The following table lists the output programming parameters for each model.

Table 4-3. Output Programming Parameters

Parameter

Value

66312A

66312A

6631B
6611C

6632B
6612C

6633B
6613C

6634B
6614C

[SOUR:]CURR[:LEV][:IMM] MAX
and
[SOUR:]CURR[:LEV]:TRIG MAX

2.0475

5.1188

10.237
5.1188

5.1188
2.0475

2.0475
1.0238

1.0238
0.5118

*RST Current Value

10% of MAX value for all models

[SOUR:]VOLT[:LEV][:IMM]MAX
and
[SOUR:]VOLT[:LEV]:TRIG MAX

20.475

20.475

8.190

20.475

51.188

102.38

*RST Voltage Value

0 V for all models

[SOUR:]VOLT:PROT[:LEV] MAX

22

22

12

22

55

110

*RST OVP Value

MAX for all models

OUTP:PROT:DEL MAX

2,147,483.647 seconds for all models

*RST Protection Delay Value

0.08 seconds

SENS:CURR:RANG

Low range = 0

20 mA for all models

High Range = 20 mA

MAX for all models

*RST Current Range

Value MAX for all models