Sense:window – Agilent Technologies 6631B User Manual

Page 54

background image

4 - Language Dictionary

54

SENSe:WINDow

This command sets the window function that is used in output measurement calculations. The following
functions can be selected:

HANNing

A signal conditioning window that reduces errors in dc and rms measurement
calculations in the presence of periodic signals such as line ripple. It also
reduces jitter when measuring successive pulse waveforms. The Hanning
window multiplies each point in the measurement sample by the function cos

4

.

Do not use the Hanning window when measuring single-shot pulse waveforms.

RECTangular

A window that returns measurement calculations without any signal
conditioning. This window may be used for pulse measurements where the
exact period of the pulse waveform is known and the measurement interval can
be set accordingly using the SENSe:SWEep:TINTerval command.

NOTE:

Neither window function alters the instantaneous voltage or current data returned in the
measurement array.

Command Syntax

SENSe:WINDow[:TYPE]

Parameters

HANNing | RECTangular

*RST Value

HANNing

Examples

SENS:WIND RECT

Query Syntax

SENSe:WINDow[:TYPE]?

Returned Parameters