Sense:function, Sense:sweep:offset:points, Sense:sweep:points – Agilent Technologies 6631B User Manual
Page 53: Sense:sweep:tinterval
Language Dictionary - 4
53
SENSe:FUNCtion
Agilent 66312A, 66332A Only
This command configures the measurement sensor to measure either voltage or current when an acquire
trigger is used. The query returns the function setting, either VOLT or CURR.
Command Syntax
SENSe:FUNCtion Parameters "VOLTage" | "CURRent" Examples SENS:FUNC "VOLT" Query Syntax SENSe:FUNCtion? Returned Parameters SENSe:SWEep:OFFSet:POINts Agilent 66312A, 66332A Only This command defines the offset in a data sweep when an acquire trigger is used. Negative values Command Syntax SENSe:SWEep:OFFSet:POINts Parameters -4095 through 2,000,000,000 *RST Value 0 Examples SENS:SWE:OFFS:POIN -2047 Query Syntax SENSe:SWEep:OFFSet:POINts? Returned Parameters Related Commands SENS:SWE:TINT SENS:SWE:POIN MEAS:ARR SENSe:SWEep:POINts This command defines the number of points in a data sweep. Command Syntax SENSe:SWEep:POINts Parameters 0 through 4096 *RST Value 2048 Examples SENS:SWE:POIN 1024 Query Syntax SENSe:SWEep:POINts? Returned Parameters Related Commands SENS:SWE:TINT SENS:SWE:OFFS MEAS:ARR SENSe:SWEep:TINTerval This command defines the time period between samples Command Syntax SENSe:SWEep:TINTerval Parameters 15.6 microseconds through 31200 seconds *RST Value 15.6 microseconds Examples SENS:SWE:TINT 31.2E-6 Query Syntax SENSe:SWEep:TINTerval? Returned Parameters Related Commands SENS:SWE:POIN SENS:SWE:OFFS MEAS:ARR
represent data samples taken prior to the trigger.