Measure:current? fetch:current, Measure:current:acdc? fetch:current:acdc, Measure:current:high? fetch:current:high – Agilent Technologies 6631B User Manual
Page 48
4 - Language Dictionary
48
MEASure:CURRent?
FETCh:CURRent?
FETCh:CURRent? applies to Agilent 66312A, 66332A Only
These queries return the dc output current.
Query Syntax
MEASure:[SCALar]:CURRent[:DC]?
FETCh:[SCALar]:CURRent[:DC]?
Parameters
None
Examples
MEAS:CURR? MEAS:CURR:DC?
Returned Parameters
Related Commands
MEAS:VOLT?
MEASure:CURRent:ACDC?
FETCh:CURRent:ACDC?
Agilent 66312A, 66332A Only
These queries return the ac+dc rms output current.
Query Syntax
MEASure:[SCALar]:CURRent:ACDC?
FETCh:[SCALar]:CURRent:ACDC?
Parameters
None
Examples
MEAS:CURR:ACDC? FETC:CURR:ACDC?
Returned Parameters
Related Commands
MEAS:VOLT:ACDC?
MEASure:CURRent:HIGH?
FETCh:CURRent:HIGH?
Agilent 66312A, 66332A Only
These queries return the High level current of a current pulse waveform. The instrument first measures
the minimum and maximum data points of the pulse waveform. It then generates a histogram of the pulse
waveform using 1024 bins between the maximum and minimum data points. The bin containing the most
data points above the 50% point is the high bin. The average of all the data points in the high bin is
returned as the High level. If no high bin contains more than 1.25% of the total number of acquired points,
then the maximum value is returned by these queries.
Query Syntax
MEASure[SCALar]:CURRent:HIGH?
FETCh[:SCALar]:CURRent:HIGH?
Parameters
None
Examples
MEAS:CURR:HIGH?FETC:CURR:HIGH?
Returned Parameters
Related Commands
MEAS:CURR:LOW? CALC:REF:HIGH