Hand Spindle Pump

CPP1000-M/-L

WIKA Operating Instructions Hand Spindle Pump

10

2) Fully open the valve, slowly and carefully. You will notice any escaping air bubbles.

3) Close the valve again.

4) Turn the star handle further in a clockwise direction, until the test specimen and/or

the reference again indicates a pressure value of approx. 50 to 100 bar.

5) Fully open the valve, slowly and carefully. If necessary, check again if any air

bubbles escape.

6) Close the valve again.

7) Turn the star handle anticlockwise all the way out.

8) Open the valve.

9) After approximately 10 seconds, close the valve.

Repeat steps 1) to 9) again, if necessary.

The CPP1000-M/-L hand spindle pump is now ready to use.

For test specimens or reference instruments with particularly large

volumes we recommend filling them with operating liquid beforehand

or using the shut-off valve for test connection, which is available as

an accessory (see chapter 9. Accessories).

3.2 Operation

The maximum permissible pressure for the CPP1000-M/-L is

1000 bar. Higher pressures may damage the pump. The refer-

ence instrument, test specimen and any connecting tubes

that are used must not be subjected to pressures above the

maximum permissible level.

We recommend unscrewing the spindle completely when you start to record

measurement values, (turning anticlockwise) to allow enough volume for measure-

ments.

To increase the test pressure turn the star handle clockwise.

To decrease the test pressure turn the star handle anticlockwise.

The use of a fine adjustment valve (CPP1000-L availabel as standard) allows to

make fine adjustments to the initial pressure setting by turning the fine adjustment

valve in or out.

The display of the instrument under test can be compared with the reference

measuring instrument at each individual calibration point.

Since small amounts of air are always compressed in the system

together with the medium, the test pressure that has been generated

first drops off slightly. After this it must be readjusted.