Gb 3. technical data – WIKA PGT43.100 User Manual
Page 8

GB
3. Technical data
3.7 Model APGT43.100 / APGT43.160
Working pressure
- Steady
Full scale range
- Fluctuating
0.9 x full scale range
- Overload capability
see data sheet PM 05.0
Temperature influence
max. ± 0.8 % of full scale range/10 K (when the temperature of the
measuring system deviates from +0 °C reference temperature)
IP protection class
- Housing
IP 54 per EN 60 59 / IEC 60 59
- With liquid filling
IP 65
Permissible temperatures
- Ambient
filled/unfilled -0 ... +60 °C
- Medium
maximum +100 °C
Further technical data see data sheet PV 15.0
3.8 Model PGT63HP.100 / PGT63HP.160
Working pressure
- Steady
Full scale range
- Fluctuating
0.9 x full scale range
- Overload capability
50 x full scale range
Temperature influence
max. ± 0.6 % of full scale range/10 K (when the temperature of the
measuring system deviates from +0 °C reference temperature)
IP protection class
- Housing
IP 54 per EN 60 59 / IEC 60 59
Permissible temperatures
- Ambient
-0 ... +60 °C
- Medium
maximum +100 °C
Further technical data see data sheet PV 16.06
3.9 Model DPGT43.100 / DPGT43.160
Working pressure
- Steady
Full scale range
- Fluctuating
0.9 x full scale range
- Overload capability
at least 10 x full scale value, max. 5 bar
Temperature influence
max. ± 0.5 % of full scale range/10 K (when the temperature of the
measuring system deviates from +0 °C reference temperature)
IP protection class
- Housing
IP 54 per EN 60 59 / IEC 60 59
- With liquid filling
IP 65
Permissible temperatures
- Ambient
filled/unfilled -0 ... +60 °C
- Medium
maximum +100 °C
Further technical data see data sheet PV 17.05
11
97
64 05/
009 GB/D/F/E
WIKA Operating Instructions intelliGAUGE family
8