beautypg.com

Specifications model wu-1x – WIKA WU-16 User Manual

Page 7

background image

2460160.06 GB/D/F/E 01/2010

12 WIKA Operating instructions/Betriebsanleitung/Mode d'emploi/Instrucciones de servicio WU-1X

2460160.06 GB/D/F/E 01/2010

13

WIKA Operating instructions/Betriebsanleitung/Mode d'emploi/Instrucciones de servicio WU-1X

Specifications

Model WU-1X

WU-10 / WU-15

WU-16

Pressure ranges

bar

4

7

10

16

25

40

60

100

160

250

400

psi

60

100

160

250

300

500

1000 1500 2000 3000 5000

Over pressure safety

1)

bar

8

14

20

32

50

80

120

200

320

500

500

Burst pressure

1)

bar

40

70

100

160

250

400

550

720

720

720

720

Other pressure ranges and pressure units (e.g. MPa, kg/cm

2

) on request

1)

1 bar = 14.50 psi

Measuring principle

Metal thin-film sensor

Materials

„

Wetted parts

2.4711 Elgiloy

®

(Sensor); 316L VIM/VAR (Pressure connection)

„

Case

Stainless steel

Surface finish

Electropolished, typical
Ra

≤ 0.18 µm (RA 7); Ra max ≤ 0.25 µm (RA 10)

Dead volume

WU-10 < 1500, WU-15 < 1000, WU-16 < 1000

Permissible Medium

Liquid / Gas / Vapour

Power supply U

B

U

B

in VDC

10 <

U

B

30 (14 ... 30 V with output signal 0.1 ... 10.1 V and 0...10 V)

Capacitive load

µF

Max. 0.1 (Voltage output)

Max. current output

mA

< 1 Voltage output; Source

mA

< 30 Current output

Adjustability zero

% of span

± 5 via potentiometer

Response time (10 ... 90 %)

ms

2

Dielectric strength

VDC

500

Accuracy

% of span

0.25

for pressure ranges > 4 bar (BFSL)

% of span

0.5

for pressure ranges

≤ 4 bar (BFSL)

% of span

0.5

2)

for pressure ranges

> 4 bar

% of span

1.0

2)

for pressure ranges

≤ 4 bar

Specifications

Model WU-1X

2)

Including non-linearity, hysteresis, zero point and full scale error (corresponds to

error of measurement per IEC 61298-2).

Adjusted in vertical mounting position with lower pressure connection.

Non-linearity

% of span

0.15 for pressure ranges > 4 bar (BFSL) according to IEC 61298-2

% of span

0.3 for pressure ranges

≤ 4 bar (BFSL) according to IEC 61298-2

1-year stability

% of span

0.2 (at reference conditions)

Permissible temperature of

„

Medium

-40 ... +100 °C

-40 ... +212 °F

„

Ambience

-20 ... +85 °C

-4 ... +185 °F

„

Storage

-40 ... +100 °C

-40 ... +212 °F

Compensated temp range

-20 ... +80 °C

-4 ... +176 °F

Temperature coefficients within

compensated temp range

„

Mean TC of zero

% of span

0.1 / 10 K

„

Mean TC of range

% of span

0.15 / 10 K

CE- conformitiy

„

Pressure equipment directive

97/23/EC

„

EMC directive

2004/108/EC, EN 61 326 Emission (Group 1, Class B) and

Immunity (industrial locations)

„

Directive ATEX of equipment

intended for use in potentially

explosive atmospheres

94/9/EC

Shock resistance

g

500 according to IEC 60068-2-27 (mechanical shock)

Vibration resistance

g

10 according to IEC 60068-2-6 (vibration under resonance)

Wiring protection

„

Short-circuit proofness

Sig+ towards U

B

- (short-time)

„

Reverse polarity protection

U

B

+ towards U

B

-

Weight

kg

Approx. 0.1

7. Starting, operation

GB

7. Starting, operation

GB

This manual is related to the following products: