WIKA WU-16 User Manual
Page 6
![background image](/manuals/465671/6/background.png)
2460160.06 GB/D/F/E 01/2010
10 WIKA Operating instructions/Betriebsanleitung/Mode d'emploi/Instrucciones de servicio WU-1X
2460160.06 GB/D/F/E 01/2010
11
WIKA Operating instructions/Betriebsanleitung/Mode d'emploi/Instrucciones de servicio WU-1X
7. Starting, operation
GB
Load (e.g. display)
Power supply
UB+/Sig+
Positive supply / measurement connection
OV/Sig-
Negative supply / measurement connection
7. Starting, operation
GB
Flying leads with 1.5 m of cable,
conducter cross section up to
max. 0.22 mm ², AWG 24
with end splices, conducter outer
diameter 4.8 mm
IP 65 - Order code: DI
2-wire
red
black
Bayonet connector
4-pin
IP 65 (NEMA 4) *
)
Order code: O4
*
)
mit Side-Access IP 54 (NEMA 3S)
SUB D connector
15-pin
IP 54 (NEMA 3S)
Order code: TX
Current for external display- or evaluation equipment can be supplied directly from the circuit,
when operating a transducer with current output. A voltage drop specific to the display unit is
to be considered. The UHP-display unit of type WUR-1 has a specific voltage drop of 6 V. The
transducers are short-circuit-proof for a short time, but anyhow any incorrect connection of
the instrument should be avoided.
Current output (2-wire)
4 ... 20 mA:
R
A
≤ (U
B
– 10 V) / 0,02 A
with R
A
in Ohm and U
B
in Volt
Voltage output (3-wire)
0.1 ... 5.1 V:
R
A
> 5 kOhm
0.1 ... 10.1 V:
R
A
> 10 kOhm
1 ... 5 V:
R
A
> 5 kOhm
0 ... 5 V:
R
A
> 5 kOhm
0 ... 10 V:
R
A
> 10 kOhm
with R
A
in Ohm and U
B
in Volt
Signal output and allowed load
4 ... 20 mA, 2-wire
permitted
range
3-wire
Circular connector
M 12x1, 4-pin
IP 65 (NEMA 4) *
)
Order code: M4
2-wire
3-wire
Circular connector R03
6-pin
IP 54 (NEMA 3S)
Order code: U6
black
red
brown