Failure

Possible cause

Procedure

Abnormal zero point signal

Medium or ambient temperature too

high/too low

Control the internal temperature of the

instrument within the permissible range;

observe the allowable temperature error

(see Operating Instructions)

Diaphragm is damaged, e.g. through

impact, abrasive/agressive media;

corrosion of diaphragm/pressure

connector.

Replace instrument

Signal span dropping off/too small

Diaphragm is damaged, e.g. through

impact, abrasive/agressive media;

corrosion of diaphragm/pressure

connector

Contact the manufacturer and replace

the instrument

Signal span too small

Power supply too high/too low

Correct the power supply in line with

the Operating Instructions

Mechanical overload through over-

pressure

Re-calibrate the instrument *)

10. Trouble shooting

GB

In case of unjustified reclamation we charge the reclamation handling expenses.
*) Make sure that after the setting the unit is working properly. In case the error continues to exist send in the instrument for

reparation (or replace the unit).

If the problem persists, contact our sales department.

USA, Canada: If the problem continues, contact WIKA or an authorized agent for assistance.
If the pressure transmitter must be returned obtain an RMA (return material authorization)
number and shipping instructions from the place of purchase. Be sure to include detailed
information about the problem. Pressure transmitters received by WIKA without a valid RMA
number will not be accepted.
Process material certificate (Contamination declaration for returned goods)
Purge / clean dismounted instruments before returning them in order to protect our
employees and the environment from any hazard caused by adherent remaining media.
Service of instruments can only take place safely when a Product Return Form has been
submitted and fully filled-in. This Return Form contains information on all materials with
which the instrument has come into contact, either through installation, test purposes,
or cleaning. You can find the Product Return Form on our internet site (www.wika.de).

11. Storage, disposal

GB

11. Storage, disposal

When storing or disposing of the transducer, take precautions with regard
to remaining media in removed transducers. We recommend cleaning the
transducer properly and carefully. Remaining media in the pressure port
may be hazardous or toxic!

Mount the protection cap when storing the transducer.

Dispose of instrument components and packaging materials in accordance with the
respective waste treatment and disposal regulations of the region or country to which
the instrument is supplied.

Storage

Disposal