NOVUS TP510 User Manual
Page 2
Legend to cross-section drawing
1
Connection fitting
2
Steel cell welded
3
Electronic with EMC-protection
4
Electrical connection (example
Quickon)
Pressure ranges
Relative pressure (Gage)
(differential measurement of pres-
sure relative to ambient pressure).
Ranges see order code selection
table
Overload
3.0x Full scale
Rupture pressure
6.0x Full scale
Accuracy
Total of linearity, hysteresis and
repeatability
< +/- 0.5% fs
Adjustment accuracy zero point and
full scale
< +/- 0.5% fs
Housing material
Casing:
Stainless steel 1.4305 (AISI 303)
Materials in contact with the medium
Sensor: Stainless steel 1.4305
Application temperature
Medium temperature
- 40 … + 150 °C
Ambient temperature with
Cable 1.5 meters
max. 85 °C
Quickon
max. 85 °C
(temperature – 150 °C on request)
Temperature influences
TC zero point < +/-
0.03% fs/K
TC sensitivity < +/-
0.015% fs/K
temperature range - 40 … + 125 °C
Dynamic response
Suitable for static and dynamic
measurements.
Response time < 2 ms
typ. 1 ms
Pressure connections
See order code selection table
Weight
Version inside thread
88 grams
Version outside thread 98 grams
Installation arrangement
Unrestricted
Outputs / Power supply
See order code selection table
• Short circuit-proof and protected
against polarity reversal. Each
connection against other with max.
+/– supply voltage.
Electric strength 500 VDC,
on request 1000 VDC
Load
0 – 5
V
> 10 kOhm < 100 nF
1 – 6
V
> 10 kOhm < 100 nF
0 – 10 V
> 10 kOhm < 100 nF
4 – 20mA
<
[Ohm]
ratiometric
>10 kOhm < 100 nF
Current consumption
With max. signal output:
0
– 5
V
<
4
mA
1
– 6
V
<
4
mA
0
–
10
V
<
4 mA
4
–
20 mA
<
20 mA
ratiometric <
4
mA
Electrical connections /
Protection standard
See order code selection table
Tests
Shock acc. IEC 68-2-27
100 G, 11 ms half sine wave, all
6 directions. Free fall from 1 m on
concrete (6x).
Constant shock acc. IEC 68-2-29
40 G for 6 ms, 1000 x all 3 direc-
tions.
Vibration acc. IEC 68-2-6
20 G, 9 ... 2000 Hz, 2 ... 9 Hz with
amplitude +/- 15 mm, 1 Octave /
minute all 3 directions, 50 con-
stant load
EMC-behaviour see on the back
UL according to standard 873
Supply voltage - 8 V
0.02 A
The distinct advantages
• Compact, rugged construction for
highest operational reliability
• Protection IP 67 standard
• Welded without sealing parts,
no elastomer-sealings
• Negligible temperature influence
on accuracy
• Excellent EMC-capacity
• Saving time by quick cable mounting
by the customer with Quickon-System
EDITION 07/2004
Technical overview
These compact OEM pressure trans-
mitters type 510 have been develo-
ped for applications in the industrial
refrigeration technology.
The special sensor is based on thick-
film technology, developed by Huba
Control, which utilises a special gra-
de of steel welded to the pressure
sensor. This means that the 510 can
be used for all gases and media in
refrigeration, including ammoniac.
Highest requirements concerning
EMC and accuracy for all tempera-
ture ranges can be met in combi-
nation with the unique integrated
electronic design.
Suitable from small series to high
quantity applications with best pri-
ce to performance ratio.