Section 7. Measurement Instructions
7-42
Parameter 
& Data Type 
Enter
CaptureTrig
Constant, 
Variable, or 
Expression 
This argument is used when the "Capture All Events" output option is used. 
When CaptureTrig is true, the INT8 will return the last nnnn events. 
Mult, Offset
Constant, 
Variable, 
Array, or 
Expression
A multiplier and offset by which to scale the raw results of the 
measurement. See the measurement description for the units of the raw 
result; a multiplier of one and an offset of 0 are necessary to output in the 
raw units. For example, the TCDiff instruction measures a thermocouple 
and outputs temperature in degrees C. A multiplier of 1.8 and an offset of 
32 will convert the temperature to degrees F. 
SDMSpeed (BitPeriod)
Changes the rate that the CR5000 uses to clock the SDM data. Slowing down 
the clock rate may be necessary when long cables lengths are used to connect 
the CR5000 and SDM devices. 
Parameter 
& Data Type 
Enter
BitPeriod
Constant or 
variable
The time per bit, in microseconds. 
Initial Setting (default): 30.0 µS 
Resolution: 50 µS 
Min Setting: 8 µS 
Max Setting 3 mS 
SDMSpeed(30) gives: 30.0 µS 
SDMSpeed(k) gives: bit_rate = INT(k*20) * Resolution 
When calculating SDMSpeed(k), the loggers round down to the next higher 
bit rate.
SDMTrigger
When SDMTrigger is executed, the CR5000 sends a "measure now" group 
trigger to all connected SDM devices. SDM stands for Synchronous Device 
for Measurement. SDM devices make measurements independently and send 
the results back to the datalogger serially. The SDMTrigger instruction allows 
the CR5000 to synchronize when the measurements are made. Subsequent 
Instructions communicate with the SDM devices to collect the measurement 
results. Not all SDM devices support the group trigger; check the manual on 
the device for more information. 
 
SIO4 (Dest, Reps, SDMAddress, Mode, Command, Param1, Param2,
ValuesPerRep, Multiplier, Offset)
The SIO4 instruction is used to control and transmit/retrieve data from a 
Campbell Scientific SIO4 Interface (4 Channel Serial Input/Output device). 
See the SDM-SIO4 Serial Input Interface manual for operation details. 