Section 7. Measurement Instructions

7-10

Parameter
& Data Type

Enter

ExChan

Enter the excitation channel number to excite the first measurement.

Constan

Alpha
Code

Code/
Channel

Result

VX1 1

Switched excitation channels, are switched to the excitation
voltage.

VX2 2

for the measurement and switched off between measurements.

VX3 3

VX4 4

MeasPEx

Constant

The number of sensors to excite with the same excitation channel before automatically
advancing to the next excitation channel. To excite all the sensors with the same
excitation channel, the number should equal the number of Reps.

ExmV

Constant

The excitation voltage in millivolts. Allowable range ± 5000 mV. RevEx may be used to
excite with both a positive and negative polarity to cancel offset voltages.

RevEx

Code Value Result (Reversing requires twice as much time to complete)

Constant

False

0

Excite only with the excitation voltage entered

True

≠0

A second measurement is made with the voltage polarity
reversed to cancel offsets

RevDiff

Code Value Result (Reversing requires twice as much time to complete)

Constant

False 0

Signal is measured with the high side referenced to the low

True

≠0

A second measurement is made after reversing the inputs to
cancel offsets

SettlingTime

Constant

The time in microseconds to delay between setting up a measurement (switching to the
channel, setting the excitation) and making the measurement. (1 microsecond resolution)

Entry

Voltage Range

Integration

Settling Time

0

± 20 mV

Flash

200 µS (default)

0

All but ± 20 mV,

Flash

100 µS (default)

0

All

250 µS

200 µS (default)

0

All

_50Hz, _60 Hz

3 mS (default)

>=100

All All

µS

entered

Integ

The time spent on integration in microseconds for each of the channels measured.

Constant

Entry

Integration

0, 200

Flash Conversion; 200 = 2 flash conversions 100 µS apart
and averaged

250; 500-16000

250 µS; multiples of 500: 250 µS integrations starting each
500 µS and averaged

_60Hz or 16667

16,667 µS (reject 60 Hz noise)

_50 Hz or 20000

20,000 µS (reject 50 Hz noise)

Mult, Offset

Constant,
Variable,
Array, or
Expression

A multiplier and offset by which to scale the raw results of the measurement.