beautypg.com

Campbell Scientific CR5000 Measurement and Control Module User Manual

Page 133

background image

Section 7. Measurement Instructions

7-7

Half Bridge Parameters

Parameter
& Data Type

Enter

Dest

Variable or
Array

The Variable in which to store the results of the instruction. When Reps are used the
results are stored in an array with the variable name. An array must be dimensioned to
have elements for all the Reps.

Reps

Constant

The number of repetitions for the measurement or instruction.

Range

Constant

Alpha Code

Numeric
Code

Voltage
Range

mV5000 0

± 5000 mV

mV1000 1

± 1000 mV

mV200 2

± 200 mV

mV50 3

± 50 mV

mV20 4

± 20 mV

AutoRange 5

mV20 - mV5000 Selects range (Sect. 3.1)

mV200C

mV50C
mV20C

20
30
40

± 200 mV

± 50 mV
± 20 mV

The mV200C, mV50C, and mV20C
ranges pull the channel into common
mode range and check for open input

SEChan

Constant

The single-ended channel number on which to make the first measurement. When Reps
are used, subsequent measurements will be automatically made on the following single-
ended channels. If the channel is entered as a negative number, all reps occur on that
channel.

ExChan

Enter the excitation channel number to excite the first measurement.

Constan

Alpha
Code

Code/
Channel

Result

VX1 1

Switched excitation channels, are switched to the excitation
voltage.

VX2 2

for the measurement and switched off between measurements.

VX3 3

VX4 4

MeasPEx

Constant

The number of sensors to excite with the same excitation channel before automatically
advancing to the next excitation channel. To excite all the sensors with the same
excitation channel, the number should equal the number of Reps.

ExmV

Constant

The excitation voltage in millivolts. Allowable range ± 5000 mV. RevEx may be used to
excite with both a positive and negative polarity to cancel offset voltages.

RevEx

Code Value

Result (Reversing requires twice as much time to complete)

Constant

False

0

Excite only with the excitation voltage entered

True

≠0

A second measurement is made with the voltage polarity
reversed to cancel offsets

RevDiff

Code Value

Result (Reversing requires twice as much time to complete)

Constant

False

0

Signal is measured with the high side referenced to the low

True

≠0

A second measurement is made after reversing the inputs to cancel
offsets