Alibration, Cal code, Stp5 stp4 stp2 – Red Lion TLA User Manual

Page 12: Stp3, Stp1, Rtd2, Rtd1

background image

Millivolt Calibration (CAL)

Connect precision millivolt source with an accuracy of 0.03% to terminals #8

(-) & #9 (+). Cold Junction or RTD ohms calibration MUST be performed after

millivolt calibration.

Thermocouple Cold Junction Calibration (CJC)

This procedure must be performed AFTER an accurate mV calibration.

1. Exit Factory Service Operations (continually press P until “End”), and return

to Normal Display Mode.

2. Connect a thermocouple probe of known accuracy to the TLA (Types T, E, J,

K, and N only). Select the probe type used in Configure Module 1.

3. Connect a reference temperature probe to the measuring end of the TLA

thermocouple probe. The two probes should be shielded from air movement

and allowed sufficient time to equalize in temperature. (As an alternative, the

TLA thermocouple probe may be placed in a calibration bath of known

temperature.)

4. Compare TLA display with reference temperature probe (or calibration bath).

If the displayed TLA temperature does not equal the reference probe

temperature, calculate the CJ error as follows:

CJ Error = reference probe temperature - displayed TLA temperature

5. Enter Factory Service Operations Module (9-FS).

RTD Ohms Calibration (RTD)

This procedure must be performed AFTER an accurate mV calibration.

Connect one leg of precision resistance (accuracy of 0.1 ohm) to terminals #9

and #10 together, and the other leg to #8.

12

6.0 C

ALIbRATION

When re-calibration is required (generally every two years), this procedure

should be performed by qualified technicians using appropriate equipment.

Equipment source accuracy of 0.03% or better is required.

The procedure consists of: applying accurate mV signals, setting the

thermocouple cold junction temperature, and applying precision resistance,

among others. Allow a 30 minute warm-up period before starting this procedure.

Do not use thermocouple wire for the millivolt or RTD ohms calibration.

This procedure may be aborted by disconnecting power to the TLA before

exiting the configuration mode. The existing calibration settings remain in

affect.

Note: After completing any of the calibration sequences, the TLA defaults the

input sensor type to thermocouple type “J” (tc-J). Be sure to set input sensor

for proper type.

Note: The TLA must be restored to normal display mode before any data is

stored.

Not required if only using TC input. This procedure can only be performed

AFTER an accurate mV calibration.

yes/no

RTD resistance calibration

rtd

Not required if only using RTD input. This procedure can only be performed

AFTER an accurate mV calibration.

yes/no

Thermocouple cold junction

temperature calibration

CJC

Calibration required for both RTD and TC input. If this procedure is performed,

the cold junction temp or RTD ohms calibration procedures in turn must be

completed.

Calibrate instrument.

yes/no

48

Millivolt calibration

Enter function code

CAL

CodE

DESCRIPTION/ COMMENTS

RANGE

PARAMETER

DISPLAY

Factory Service Operations - Calibration (9-FS)

Wait 10 seconds, press P.

Apply 14.0 mV

Wait 10 seconds, press P.

Wait 10 seconds, press P.

Apply 56.0 mV

Apply 42.0 mV

StP5

StP4

StP2

Wait 10 seconds, press P.

Apply 28.0 mV

StP3

Apply 0.0 mV

StP1

DESCRIPTION/ COMMENTS

ACTION

DISPLAY

Wait 10 seconds, press P.

Cold Junction

Temperature

CJC

DESCRIPTION/ COMMENTS

PARAMETER

DISPLAY

Note: If the initial value for CJC is not within

the range of 15°C to 40°C, enter 25.0° for

CJC and repeat the Cold Junction Calibration

procedure.

Observe the indicated cold junction

temperature. Add the calculated CJ Error to

the displayed value. Enter the sum as the new

value for CJC. Exit 9-FS and repeat step 4.

Wait 10 seconds, press P.

rtd2

Wait 10 seconds, press P.

rtd1

DESCRIPTION/ COMMENTS

ACTION

DISPLAY

Connect 0.0 ohm

(jumper wire)

Connect 277.0 ohm