beautypg.com

Vaisala GM70 User Manual

Page 62

background image

USER'S GUIDE ______________________________________________________

62

_____________________________________________________ M010139EN-G

Table 6

Accuracy Specifications

Accuracy (including repeatability, non-linearity and calibration
uncertainty) at 25°C and 1013 hPa.

Probe Accuracy
GMP221

Accuracy specification applies for
CO

2

concentrations above 2% of

full scale

±[1.5% of range + 2% of reading]

GMP222

±[1.5% of range + 2% of reading]

Property

Description / Value

Temperature
dependence, typical

-0.3% of reading / °C (reference
25ºC/77ºF)

Pressure dependence,
typical

+0.15% of reading / hPa (reference
1013 hPa)

Long-term instability

<±5%FS / 2 years

Response time (63%) 20 seconds (GMP221)

30 seconds (GMP222)

Table 7

Other Specifications

Property

Description / Value

Warm-up time

with GM70PUMP

30 seconds
15 min full specification
30 min full specification

Operation temperature -20 … + 60ºC
Operation pressure

700 … 1300 hPa

Operation humidity

0 … 100%RH non-condensing

Flow range (diffusion
sampling)

0 … 10 m/s

Probe material

PC

Probe cable length

1.9 m