1.

Insert the probe into the field check adapter until the perforated
filter is covered.

0505-237

Figure 17

Field Check Adapter

2.

Connect the tubing to the bottom port of the adapter.

3.

Connect the adapter with the tubing to the flow meter, the pressure
regulator and further to the low-end reference gas bottle. The side
port of the adapter is left open for gas outflow.

4.

Let the low end reference gas flow and stabilize with a flow rate of
about 0.6 l/min. Follow the reading of GM70, and after it has
stabilized, wait for 6 more minutes.

5.

Take out the screw from the GM70 probe handle to expose the
adjustment button (see Figure 16 on page 48). Press the button with
a small screwdriver. When pressing the button, the indicator turns
to adjusting mode.

6. Press

OK

to confirm the adjusting.

7. Select

CO2

, press

SELECT.

8. Press

YES

to give the ambient pressure and temperature values.

To continue adjusting press

EXIT

.

9.

Now the adjustment mode is on. Press

GRAPH

to confirm that the

readings have stabilized. Go back and press

ADJUST

to select the

adjustment method.

10. Select

2-point adjustment

, press

SELECT

. Press

READY

with the

stabilized reading in the lower reference concentration.

11. Give the lower reference concentration value by using the arrow

buttons (for example, if you are using pure nitrogen, enter value 0
ppm). Take care that you give the correct value in correct units, as
you cannot revert the earlier values after accepting the new values.
Press

OK

.

Next move on to the adjustment at the second (high-end)
reference point.

12. Take out the tubing from the low-end gas bottle and connect it to

the high-end gas bottle.