Vaisala GM70 User Manual

Page 56

background image

USER'S GUIDE ______________________________________________________

56

_____________________________________________________ M010139EN-G

Error Message

Interpretation and Action
power.

Probe error

This error can occur when changing the probe while power is on.
Switch off power and check that the probe is connected properly:
unscrew the probe fastener. Insert the probe to the handle as
deep as possible. Turn the probe inside the handle until you feel
that a step in the probe connector snaps into the groove of the
handle connector and locks the probe. Tighten the probe
fastener. Switch on power.

Adapter error

Switch off power and check that the probe is connected properly,
see the instructions of Probe error. In case of continuous error,
contact Vaisala or send the probe to Vaisala Service, see
sections Technical Support and Product Returns on page 60.