View recorded data – Vaisala PTB330 User Manual
Page 93

Chapter 4 ________________________________________________________________ Operation
VAISALA _______________________________________________________________________ 91
Example:
>dsel p HCP
P HCP
>
Enter the command without parameters and press ENTER to display
current recording parameters.
NOTE
If P
3h
is selected, P is recorded instead.
View Recorded Data
If the device is provided with the optional display, the graphical
display shows the data of the selected quantities, one at a time. See
section Graphic History on page 51 for details about graphical display.
You can also dump the logged data to the serial line in numeric form
with the commands presented below.
DIR
Use the serial line and enter the DIR command to list the logged data
files.
The device records six files (six observation periods) for each selected
quantity. Thus, total amount of the files depends on the amount of the
selected quantities being at minimum 6 and at maximum 24. See
Table 8 on page 52.
NOTE
The timestamps provided by the DIR command are based on the
current
system date and time. In order to see correct timestamps, you
must ensure that the system date and time is set to correct values with
DATE
and TIME commands (see section DATE and TIME on page
75) before using the DIR command.