Measurement related commands, Tqfe – Vaisala PTB330 User Manual
Page 79

Chapter 4 ________________________________________________________________ Operation
VAISALA _______________________________________________________________________ 77
Measurement Related Commands
TQFE
Use the TQFE command to show or set the temperature for QFE
corrected pressure. The valid range of QFE temperature is -80...+200
°C.
TQFE [?] [x] [y]
where:
? = Displays the current QFE temperature. You cannot use this
modifier with the other modifiers.
x = Numeric value (temperature).
Ranges:
-80 … 200 °C
-110…390 °F
190…470 K
y = Temperature scale used.
Values:
‘C
‘F
K
Examples:
Query syntax:
>TQFE ?
QFE temp. : 20.00 'C
>
Prompt syntax:
>TQFE
QFE temp. : 20.00 'C ? 70 'F
>
Command line syntax:
>TQFE 300 K
QFE temp. : 300.00 K
>
See also other documents in the category Vaisala Tools:
- DM500 (138 pages)
- DM70 (93 pages)
- DMT132 (74 pages)
- DMT143 (76 pages)
- DMT152 (70 pages)
- DMT242 (4 pages)
- DMT340 (191 pages)
- DMT345 (185 pages)
- DPT145 (63 pages)
- DPT146 (71 pages)
- PTU300 (217 pages)
- PTB330TS (89 pages)
- PTB220 (10 pages)
- PTB220 (113 pages)
- PTU200 (64 pages)
- PTU200MIK1 (18 pages)
- SPH10 (2 pages)
- SPH20 (2 pages)
- PTB110 (4 pages)
- PTB200 (30 pages)
- PTB210 (analog) (27 pages)
- PTB210 (serial) (32 pages)
- GM70 (68 pages)
- GMD20 (4 pages)
- GMK220 (18 pages)
- GML20 (2 pages)
- GML20T (2 pages)
- GMM20W (5 pages)
- GMM220 (6 pages)
- GMP231 (2 pages)
- GMP231 (90 pages)
- GMP343 (94 pages)
- GMR20 (2 pages)
- GMT220 (42 pages)
- GMW90 (101 pages)
- XMW90 (4 pages)
- MM70 (67 pages)
- MM70 (71 pages)
- MMT162 (66 pages)
- MMT310 (81 pages)
- MMT330 (181 pages)
- MMT330 (171 pages)