Display settings, Changing quantities, Changing units – Vaisala PTB330 User Manual
Page 56

USER'S GUIDE____________________________________________________________________
54 __________________________________________________________________ M210855EN-D
Display Settings
Changing Quantities
1. Open
the
MAIN MENU by pressing any of the arrow buttons.
2. Select
Display and press the right arrow button.
3. Select
Quantities and press the right arrow button.
4.
Select a quantity by using the up/down arrow buttons. Confirm
your choice by pressing the SELECT button. You can select 1
... 4 display quantities at a time. Undo selection by selecting it
and pressing the HIDE button.
5. Press
EXIT to return to the basic display.
See Table 2 on page 15 for available quantities.
NOTE
Only the quantities selected for the display are stored in the graphical
history. If a quantity is hidden, it is not stored in the memory of the
barometer for the time being.
NOTE
In order to output P
3h
or A
3h
readings through the serial line, either P
or P
3h
must
be selected for display.
Changing Units
1. Open
the
MAIN MENU by pressing any of the arrow buttons.
2. Select
Display and press the right arrow button.
3.
Use the up/down arrow buttons to select Units. Press the right
arrow button.
4.
Use the up/down arrow buttons to select display units. Confirm
the selection by pressing CHANGE.
5. Press
EXIT to return to the basic display.
NOTE
Changing the display quantities/units (by using the display/keypad)