Opening and closing the adjustment mode – Vaisala PTB330 User Manual
Page 114
![background image](/manuals/408510/114/background.png)
USER'S GUIDE____________________________________________________________________
112 _________________________________________________________________ M210855EN-D
Table 19
Adjustment and Calibration Commands for
Barometer Module P1
Command Function
LCP1 ?
Displays current linear adjustment
status
LCP1 ON/OFF
Enables/Disables the linear
adjustments
LCP1
Enter linear adjustments
MPCP1 ON/OFF
Enables/Disables the multipoint
adjustments
MPCP1
Enters multipoint adjustments
CDATE
Shows or sets the calibration date
CTEXT
Shows or sets the calibration info
text.
Opening and Closing the Adjustment Mode
Before you can proceed with the adjustment and calibration, you must
open the adjustment mode by following the instructions below. The
adjustment button with indicator LED is located in the upper corner
on the left-hand side of the motherboard. See the position of the
adjustment button in Figure 2 on page 16. To open and close the
adjustment mode:
1.
Open the barometer cover.
2. Press
the
ADJ
button. The adjustment mode is now available.
The red indicator LED is on.
3. Press
the
ADJ
button again to lock the adjustment mode. The
red indicator LED is off.
The adjustment menu on the display/keypad is visible only when the
ADJ
button (on the motherboard inside the transmitter) is pressed.
Figure 44
Adjustment Menu