Chapter 4 ________________________________________________________________ Operation
VAISALA _______________________________________________________________________ 69
Measurement setting commands
TQFE
[-80...+200° C]
Shows or sets the temperature for QFE
corrected pressure.
DPMAX
[0…9999.99 hPa] Shows or sets the maximum permissible
difference pressure between barometer
modules.
HHCP
[ -30...+30 m]
Shows or sets the altitude for height
corrected pressure.
HQFE
[ -30...+30 m]
Shows or sets the altitude for the QFE
corrected pressure.
HQNH
[ -30...+3000 m]
Shows or sets the altitude for the QNH
corrected pressure.
PSTAB
[0…9999.99 hPa] Shows or sets the pressure stability limits.
AVRG
[0…600 s]
Sets the barometer measurement
averaging time (in seconds).
Formatting commands
FORM
Sets the custom output for the SEND
command and for RUN mode.
TIME
Shows or changes the current time.
DATE
Shows
or
changes
the current date.
UNIT
Shows or sets unit for a quantities.
Data recording commands
DSEL
Selects the quantities that are displayed on
the graphical user interface as well as
quantities for data recording.
DELETE
Erases the log memory.
UNDELETE
Restores the erased log memory.
DIR
Lists the available logs in the logging
memory.
PLAY
Shows the trend, min, and max values of
the given log.
Calibration and adjustment commands
CDATE
Shows or sets the calibration date.
LCP1/
LCP2/
LCP3
Performs a linear adjustment for the
barometer module/module.
MPCP1/M
PCP2/MP
CP3
Performs a multipoint adjustment for the
barometer module/module.
CTEXT
Shows or sets the calibration info text.
NOTE
The calibration and adjustment commands are available in adjustment
mode only. Press the adjustment button before inputting these
commands.