USER'S GUIDE____________________________________________________________________
88 __________________________________________________________________ M210855EN-D
NOTE
Because ERRS command does not display any difference between
present and past errors, execute ERRS twice in a row to see which
ones are still present (past errors are not displayed anymore on the
second ERRS response).
Table 17
ERRS command line parameters
Value Description
Display currently active and past, unviewed (unacknowledged)
errors.
LOG
Show error log contents.
BARO
Show detailed error information for pressure transducer errors.
Example:
>errs
PASS
No errors
>errs
FAIL
Error: Operating voltage out of range.
>
VERS
Use the command VERS to display software version of PTB330:
>VERS
MT300 / 1.13
>
Resetting Barometer by Using
Serial Line
RESET
Resets the device. The user port switches to start-up output mode
selected with command SMODE.