Set transmitter address – Vaisala DMT143 User Manual
Page 56

USER'S GUIDE ____________________________________________________________________
54 ___________________________________________________________________ M211435EN-E
SERI [b p d s]
where
b
=
baud rate (4800, 9600, or 19200)
p
=
parity (n = none, e = even, o = odd)
d
=
data bits (7 or 8)
s
=
stop bits (1 or 2)
Example (set serial line to 9600 baud, no parity, 7 data bits, and 1 stop
bit):
seri 9600 e 7 1
SCI Baud P D S: 9600 E 7 1
Set Transmitter Address
To operate in the POLL mode, the transmitter must have an address. If
multiple transmitters share the same serial line, each transmitter must
have a different address.
For a description of the serial interface modes, see section Set Serial Line
Operating Mode on page 53.
ADDR [address]
where
address
=
Transmitter address, range 0 ... 255.
Example:
addr
Address : 0 ?
- DM500 (138 pages)
- DM70 (93 pages)
- DMT132 (74 pages)
- DMT152 (70 pages)
- DMT242 (4 pages)
- DMT340 (191 pages)
- DMT345 (185 pages)
- DPT145 (63 pages)
- DPT146 (71 pages)
- PTU300 (217 pages)
- PTB330TS (89 pages)
- PTB220 (10 pages)
- PTB220 (113 pages)
- PTB330 (144 pages)
- PTU200 (64 pages)
- PTU200MIK1 (18 pages)
- SPH10 (2 pages)
- SPH20 (2 pages)
- PTB110 (4 pages)
- PTB200 (30 pages)
- PTB210 (analog) (27 pages)
- PTB210 (serial) (32 pages)
- GM70 (68 pages)
- GMD20 (4 pages)
- GMK220 (18 pages)
- GML20 (2 pages)
- GML20T (2 pages)
- GMM20W (5 pages)
- GMM220 (6 pages)
- GMP231 (2 pages)
- GMP231 (90 pages)
- GMP343 (94 pages)
- GMR20 (2 pages)
- GMT220 (42 pages)
- GMW90 (101 pages)
- XMW90 (4 pages)
- MM70 (67 pages)
- MM70 (71 pages)
- MMT162 (66 pages)
- MMT310 (81 pages)
- MMT330 (181 pages)
- MMT330 (171 pages)