Extend Analog Output Range

Use the AOVER command to allow the analog output channels to
exceed their specified range by 10%. The scaling of the parameter
remains as before; the extra range is used for additional measurement
range in the high end of the scale.

AOVER [ON/OFF]

Example (turning AOVER on):

aover on
AOVER : ON

The following example illustrates how the analog output is affected. The
channel outputs T

d

with voltage output 0 … 5 V (-80 … -30 °C). After

giving the AOVER ON command, the range is 0 … 5.5 V
(-80 … -25 °C). Note that the -30 °C T

d

point is still at 5 V.

Test Analog Output

You can test the operation of the analog output with the ATEST
command by forcing the output to a given value. You can then measure
the output with a calibrated multimeter. After testing the output, give the
ATEST command again to exit the test mode.

ATEST [level]

where

level = Level for analog output channel.

Analog output type determines if the value is V or mA.


The command output shows the test value of the analog output as well as
diagnostic information that may be useful to Vaisala Service if there is a
problem with the analog output.

Example (enabling analog output test mode, setting level to 20 mA):

atest 20
20.000 27407

Example (disabling analog output test mode, resuming normal output):

atest
0.000 15831