Chapter 5 _________________________________________________________________ Operation

VAISALA ________________________________________________________________________ 51

Select Unit

Use the UNIT command to select metric or non-metric output units.

UNIT [x]

where

x

=

Selects the unit type to output:
m = metric units, for example, Celsius
n = nonmetric units, for example, Fahrenheit


Example:

unit m
Units : Metric

View User Adjustment Parameters

Use the L command to view the current user adjustment parameters that
are affected by the LI command and the MI70 (Tdf offset only). This
command is useful for checking the currently applied correction.

L

The output values are as follows:

- RH offset and gain

- T offset and gain

- Tdf offset and gain

Example (shows default values, no user adjustment done):

l
RH offset : 0.00000000E+00
RH gain : 1.00000000E+00
T offset : 0.00000000E+00
T gain : 1.00000000E+00
Tdf offset: 0.00000000E+00
Tdf gain : 1.00000000E+00

Example (shows that a -1.2 °C Tdf offset has been applied by the user):

l
RH offset : 0.00000000E+00
RH gain : 1.00000000E+00
T offset : 0.00000000E+00
T gain : 1.00000000E+00
Tdf offset: -1.20000000E+00
Tdf gain : 1.00000000E+00