List of serial commands, Table 7 – Vaisala DMT143 User Manual
Page 40

USER'S GUIDE ____________________________________________________________________
38 ___________________________________________________________________ M211435EN-E
List of Serial Commands
All commands can be issued either in uppercase or lowercase. In the
command examples, the keyboard input by the user is in bold type.
The notation
your computer keyboard. Enter a
before starting to enter commands.
Table 7
List of Serial Commands
Command
Description
Stop the continuous outputting of RUN mode
or R command
?
Show device information
??
Show device information (overrides poll mode)
ADDR [0 … 255]
Show or set transmitter address
AERR
Set analog output error level
ALARM
Set dewpoint level for LED alarm
AMODE
Show analog output mode
AOUT [ON/OFF]
Enable or disable analog output
AOVER [ON/OFF]
Allow analog outputs to exceed their range 10%
ASEL
Set analog output parameter and scaling
ATEST
Test analog output
CLOSE
Close the temporary connection and return the
device to POLL mode
ERRS
Display the currently active errors
FILT [0 ... 1]
Set output filtering level
FORM [modifier string]
Set output format
FRESTORE
Restore factory settings
GAS [0 ... 1]
Set measured gas type
HELP
Show list of currently available serial commands
INTV [0 ... 255 S/MIN/H] Set the continuous output interval
(for RUN mode and R command)
L
Show user adjustment values
LI
Set user adjustment values
OPEN [0 … 255]
Open a line to transmitter at defined address
PRES [0.001 ... 50]
Set permanent pressure compensation value
R
Start the continuous outputting
RESET
Reset the transmitter
SDELAY [0 … 255]
Set response sending delay
SEND [0 … 255]
Output readings once
(specify address for transmitters in POLL mode)
SERI [baud p d s]
Show or set the serial interface settings
SMODE [mode]
Show or set startup serial mode:
RUN, STOP, or POLL
SNUM
Display transmitter serial number
SYSTEM
Show transmitter firmware information
TIME
Show transmitter uptime
UNIT [m/n]
Select metric or non-metric units
VERS
Show transmitter firmware version
XPRES [0.001 ... 50]
Set temporary pressure compensation value