Enable or disable analog output, Serial line output commands, Start measurement output – Vaisala DMT143 User Manual
Page 47: Stop measurement output

Chapter 5 _________________________________________________________________ Operation
VAISALA ________________________________________________________________________ 45
Enable or Disable Analog Output
Use the AOUT command to enable or disable the analog output channel.
AOUT [ON/OFF]
Example (turn analog output off):
aout off
AOUT : OFF
Serial Line Output Commands
Start Measurement Output
Use the R command to start the continuous outputting of measurement
values as an ASCII text string to the serial line. The format of the
measurement message is set with the FORM command.
R
Example (measurement message in default format):
r
Tdf= 17.75 'F Tdfatm= -5.25 'F H2O= 954 ppm
Tdf= 17.23 'F Tdfatm= -5.72 'F H2O= 930 ppm
Tdf= 17.06 'F Tdfatm= -5.88 'F H2O= 922 ppm
Tdf= 16.97 'F Tdfatm= -5.96 'F H2O= 918 ppm
...
Outputting the results continues in intervals issued with the command
INTV until stopped.
Stop Measurement Output
You can stop the measurement output by pressing the ESC key.