Set Output Interval

Use the INTV command to change the output interval of the
automatically repeating measurement messages. The measurement
messages are repeated in the RUN mode, or after the R command has
been given.

INTV [n xxx]

where

n

=

time interval, range 0 ... 255

xxx

=

time unit = "S", "MIN", or "H"


The shortest output interval (with n = 0) outputs the measurement
messages as quickly as the transmitter’s internal measurement cycle
produces them, without additional delay.

Example:

intv 1 min
Output interval: 1 MIN

Output a Reading Once

Use the SEND command to output a single measurement message. The
output message is sent in the format that has been set using the FORM
command.

SEND [address]

where

address =

Address of the transmitter, range 0 ... 255.

Specifying the address is only necessary if the target
transmitter is in POLL mode, and has not been accessed
with the OPEN command.


Example:

send
Tdf= 16.71 'F Tdfatm= -6.19 'F H2O= 907 ppm