Specifications

Table 1: Humicap Relative Humidity Measurement Specifications

(with HMP110)

Property

Description/Value

Measurement range

0 ... 100% RH

Accuracy (including non-linearity, hystere-

sis, and repeatability):

at 0 ... +40 °C (+32 ... +104 °F)


at -40 ... 0 °C and +40 ... +80 °C

(-40 ... +32 °F and +104 ... +176 °F)

±

1.7 %RH (0 ...90 %RH)

±

2.5 %RH (90 ... 100 %RH)

±

3.0 %RH (0 ...90 %RH)

±

4.0 %RH (90 ... 100 %RH)

Factory calibration uncertainty at 20 °C

(+68 °F)

±

1.5 %RH

Humidity sensor

HUMICAP 180R

Stability

±

2 %RH over 2 years

Table 2: Humicap Temperature Measurement Specifications

(with HMP110)

Property

Description / Value

Measurement range

-40 ... +80

°

C (-40 ... +176

°

F)

Accuracy over temperature range:

at +15 ... +25 °C (+59 ... +77 °F)

±

0.2 °C (

±

0.36 °F)

at 0 ... +15 °C and +25 ... +40 °C

(+32 ... +59 °F and +77 ... +104 °F)

±

0.25 °C (

±

0.45 °F)