beautypg.com

Vaisala HMT140 User Manual

Page 25

background image

Mounting

M21488EN-C

17

The following numbers refer to Figure 4:

1

Probe (HMP110)

2

Duct installation kit

3

Probe cable

To assemble the probe:
1

Pass the probe cable through the plastic pipe of the duct

installation kit.

2

Connect the probe cable to the HMP110.

Figure 5: Drilling Instructions

The following numbers refer to Figure 5:

1

Mounting screw

2

Tension screw

3

HMP110 assembled in duct installation kit plastic pipe

To attach the probe assembly to the duct:
1

Drill the holes for the duct installation kit as follows:

• Use a 24-mm drill bit to drill a hole in the wall.
• Use a 3.2-mm drill bit to make four holes for the

installation screws (four ST4.2×16-C-Z DIN 7981

screws). The holes should be arranged in a square

around the 24 mm hole, at a distance of 42 mm from

each other.

2

Mount the probe holder using the screws.

3

Adjust the depth of the plastic pipe and tighten the screw

to lock the probe in place.

42

1

2

3

Ш24

Ш3.

2