Battery replacement – Vaisala HMT140 User Manual
Page 57
Battery Replacement
M211488EN-C
49
For calibration of HMP110 probes, refer to the supporting
documentation available on the Vaisala website.
Battery Replacement
To replace the batteries and reset the battery meter of the
HMT140:
1
Open the case. Use your thumb to press and hold the
release tab, located between the two glands, then pull the
cover out and up with your fingers. See HMT140
Components Diagram on page 19 for location of release
tab.
2
Turn the power switch off.
3
Remove the batteries by pulling them up firmly. If the
battery does not come out easily, gently pull back on one
of the battery holder tabs first.
Note:
Use only Vaisala supplied batteries,
part #: 236318SP Vaisala 3.6V Lithium
Thionyl Chloride batteries.
4
After replacing the batteries, press and hold the
SERVICE button and turn the power switch on. The LED
light will flash and the device will beep eight times.
5
Release the SERVICE button. At this point, the battery
meter is also reset.
CAUTION:Remove batteries before shipping