beautypg.com

Vaisala HMT100 User Manual

Page 46

background image

User's Guide ______________________________________________________________________

46 __________________________________________________________________ M210701EN-D

0505-247

Figure 18

Accuracy of Temperature Measurement

Table 7

Dewpoint Temperature (Calculated) Measurement
Specifications

Property

Description / Value

Measuring range

-20 ... +80 °C (+4 ... 176 °F)

Accuracy see

Table 8 below

Table 8

Accuracy of Dewpoint Temperature Measurement

Relative

humidity

Temp.

10 20 30 40 50 60 70 80 90 100

-20

2.6 1.7 1.4 1.2 1.1 1.1 1.0 1.0

-10

2.7 1.7 1.3 1.2 1.1 1.0 1.0 1.0 0.9

0

2.3 1.4 1.1 1.0 0.9 0.8 0.8 0.8 0.7 0.7

10

2.3 1.4 1.1 0.9 0.8 0.9 0.8 0.8 0.8 0.8

20

2.2 1.2 1.0 0.8 0.7 0.6 0.6 0.5 0.5 0.6

40

3.0 1.9 1.5 1.3 1.2 1.1 1.0 1.0 1.0 1.0

60

4.3 2.7 2.2 1.9 1.8 1.7 1.6 1.5 1.5 1.5

80

4.9 3.1 2.5 2.2 2.1 2.0 1.9 1.8 1.8 1.8

Table 9

Operating Environment Specifications

Property

Description / Value

Operating temperature range

Transmitter body, no display
Transmitter body, with display
Sensor head


-40 ... +60 °C (-40 ... +140 °F)
-30 ... +60 °C (-22 ... +140 °F)
-40 ... +80 °C (-40 ... +176 °F)

Storage temperature range

-40 ... +60 °C (-40 ... +140 °F)

Electromagnetic compatibility

EN61326-1:1197+Am1:1998; +
Am2:2001; Generic Environment