Chapter 6, Technical data, Specifications – Vaisala HMT100 User Manual
Page 45

Chapter 6 ____________________________________________________________ Technical Data
VAISALA _______________________________________________________________________ 45
CHAPTER 6
TECHNICAL DATA
This chapter provides the technical data of the product.
Specifications
Table 5
Relative Humidity Measurement Specifications
Property
Description / Value
Measurement range
0 ... 100 % RH
Accuracy against factory standards
including non-linearity, hysteresis,
and repeatability
at +15 ... +25 °C
±1.7 %RH (0 ... 90 % RH)
±2.5 %RH (90 ... 100 % RH)
at 0 ... +40 °C
±(1.7 + 0.015 × reading) % RH
at -40 ... 0 °C and +40 ... +80 °C
±(2.0 + 0.025 × reading) % RH
Factory calibration uncertainty at
20 °C (68 °F)
±1.0 %RH (0 ... 15 % RH)
±1.5 %RH (15 ... 78 % RH)
Response time (90 %) at 20 °C in
still air
8 s with plastic grid
20 s with membrane filter
40 s with sintered filter
Humidity sensor
HUMICAP
®
180
Table 6
Temperature Measurement Specifications
Property
Description / Value
Measurement range
-40 ... +80 °C (-40 ... 176 °F)
Accuracy at +20 °C (+68 °F)
±0.2 °C (See Figure 18 on page 46)
Temperature sensor
Pt1000 IEC 751
⅓ class B